Search Website more Flowcells more Spectrophotometer Cells

Starna Flow Cells, Type 585.1 series

Ultramicro flow cell, 1mm diameter chamber
M6 threaded connectors


Description: Starna Spectrophotometer Ultramicro Flow Cell
Physical Configuration: Flowcell with the smallest volume available. The sample chamber is a hold drilled through the black cell body.
Primary use: Absorption measurement in useable range (see charts below)
Clear windows: Two
Instrument: Designed for use in spectrophotometers which can use cells with tiny aperatures
Path lengths available: 10 mm
Aperture dimensions: 1 mm diameter
Available 'Z' dimensions: 8.5, 15 and 20mm, check the table below for the "Z" dimension of your instrument
Usage note: The aperture is very tiny in this cell. Many spectrophotometers have light beams which may not be able to pass light through such a small aperture.
Connections: M6 threaded connectors. These connectors screw into the top of the cell and come with PTFE tubing which is 1.7mm OD, 0.7mm ID and about 400mm long.
"Z" Dimensions of Instrument Manufacturers
Instrument Manufacturer Z Dimension in millimeters Instrument Manufacturer Z Dimension in millimeters
Agilent® 15 mm Perkin-Elmer® 15 mm
Beckman® 8.5 mm Pharmacia® 15 mm
Bio-Rad® 8.5 mm Shimadzu® 15 mm
Eppendorf® 8.5 mm Spectronics® 8.5 mm
Hewlett Packard® 15 mm Turner® 8.5 mm
Hitachi® 8.5 mm Varian® 20 mm

Window material: Spectrosil Quartz or equivalent, useable range: 170 to 2700 nm
Starna Path Exterior Interior 'Z' Nominal Number
Catalog Length Width Length Height Diameter Dimension Volume Polished
Number mm mm mm mm mm mm ml Windows
585.1/Q/10/Z8.5 10 12.5 12.5 35 1 8.5 0.008 2
585.1/Q/10/Z15 10 12.5 12.5 35 1 15 0.008 2
585.1/Q/10/Z20 10 12.5 12.5 35 1 20 0.008 2

Starna Cells, Inc PO Box 1919 Atascadero, CA 93423 USA
Phone: 805-466-8855 Fax: 805-461-1575 e-mail: sales@starnacells.com